Laminar Flow Units

Laminar Flow Units For Modular Operation theatre

HECTRA Range of Laminar Flow Units are good quality in construction, low level of operating noise and optimum product protection. The unit is easily disassembled and assembles for free movement from one place to other. Air is drawn through a H14 type HEPA filter and blown in a very smooth. LAFU’s are UV lamps protection for sterilize the interior and contents when not in use. LAFU’s are used at medical, pharmaceutical, electronic and industrial sectors. Laminar flow is only for product protection not for personal protection. Maghenelic Guage/ digital alarm type HEPA filter status arrangement. for more details click here

Horizontal Laminar Airflow

Laminar Flow Units
Hectra Horizontal Laminar Airflow Hoods offer a HEPA filtered air is passed through the main chamber of the clean work bench in a horizontal laminar air stream and exhausted through from opening. It is ideal for operations requiring a particle-free, bacteria-free, clean air atmosphere. The clean work area provides an excellent work space for small laboratory utilization, microscopes, pipetting, IV admixture preparation, or any application requiring work-in-progress progress from outside potentially harmful contaminants.

Technical Specification

Direction of flow: Horizontal
Cleanliness: Class 100 as per US FED STD 209E (IS 14644-1)
Velocity: 90FPM±20
Noise Level: 65db max
Power Supply: 230V, 1Ø, 50Hz
Particle Retention: 0.3 micron
Supply filter: Minipleat type HEPA Filter efficiency 99.97% Down to 0.3µ H-13 Rating
Pre Filter: Micro glass fibre media of EU-4 Rating
Motor Blower Assembly: Centrifugal Blower statically and dynamically balanced with Suitable Rating

Hectra Model Number SHL 222 SHL 322 SHL 422 SHL 622
Working Size (W x D x H)Ft 2’x2’x2’ 3’x2’x2’ 4’x2’x2’ 6’x2’x2’

Request A Quote